[ English | Bahasa Malaysia ] Hari ini ialah 3 May 2024, 04:38 AM (Terakhir dikemaskini pada: 2nd May 2024)

Text Red Text Black Text Blue Text Green Text Smaller Text Reset Text Larger

Publication Details
Year :

2013

Journal :

Ruziana Mona Wan Mohd Zin, Zatil Iwani Ahmad Kamil, Tuan Rosidah Tuan Soh, Mustaffa Embong and Wan Nazaimoon Wan Mohamud (2013). Haemaglobin A1c: comparing performance of two point of care devices with laboratory analyser. BMC Research Notes 6: 540

Abstract :

Background: Measurement of HbA1c has been widely used for long-term monitoring and management of diabetes control. There is increasing use of point-of-care (POC) devices for measuring HbA1c where quicker results would allow immediate clinical management decisions to be made. Therefore, it is important to evaluate and compare the performance of such devices to the reference laboratory method.

Findings: A total of 274 venous blood was collected from normal healthy adults during the community screening programmes. The performance of POC devices, Afinion and Quo-test were compared to central laboratory HPLC method; Adams A1c HA 8160. Both POC devices showed good correlation to HA 8160 with r = 0.94 (p < 0.001) and r = 0.95 (p < 0.001) for Afinion and Quo-test respectively. The means difference were statistically higher between POC and HA 8160 with 0.23% (95% CI 0.19-0.26, p < 0.001) and 0.29% (95% CI 0.24-0.34, p < 0.001) for Afinion and Quo-test respectively.

Conclusions: Both POC devices could be considered in health clinics for diabetes management but not to be used for the diagnostic purposes.

myGOV MAMPU JPA IKU MSC NMMR CRC MyHEALTH MOH Globinmed SPA krste.my

Notis

 

Laman ini sesuai dilihat menggunakan Firefox (24.0) dengan resolusi 1280 x 768 pixels atau lebih.

Lokasi Kami

world

Kami terletak di Jalan Pahang, 50588 Kuala Lumpur, Malaysia. Klik pada imej peta untuk melancarkan servis Google Maps.

Hubungi Kami